Skip to content

Public FDA record

Z-0166-2020: Rifton E-Pacer, SKU K660, powered lift device

Recall record Z-0166-2020 for Community Products, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0166-2020
Company in FDA record
Community Products, LLC
FDA record creation date
Oct 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A design change to the load-bearing quick release buckle renders the springs in the latch mechanism susceptible to fatigue failure after repeated use.

Source facts

Action
Urgent Medical Device Correction emails were sent to customers beginning on March 21, 2019. Customers were contacted via telephone beginning April 5, 2019.
Initiated
2019-03-21
Root cause
Nonconforming Material/Component
Terminated
2020-06-15
Distribution
The products were distributed US nationwide. The products were distributed to the following foreign countries: Foreign Countries: Worldwide, including Canada, United Kingdom, Germany, Japan, Singapore, South Korea, United Arab Emirates, Taiwan, Denmark, Italy, Hong Kong, Ireland, Australia, New Zealand, Israel, Colombia, Russia
Recall number
Z-0166-2020
Recall status
Terminated
CDRH record ID
172029
FDA record created
2019-10-22
Recall event number
82607

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗