Skip to content

Public FDA record

Z-0167-2013: SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S

Recall record Z-0167-2013 for Sorin Group USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0167-2013
Company in FDA record
Sorin Group USA, Inc.
FDA classification date
Oct 29, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.

Source facts

Action
The firm, Sorin Group USA, Inc., sent an "URGENT FIELD SAFETY NOTICE" dated September 20, 2012 to all affected customers by certified mail. The notice described the product, problem and actions to be taken. Customers were instructed to discontinue use of and segregate affected product for return and replacement to the firm; to complete and return the attached Customer Response Form via fax to: 303-467-6502 or email to: yvonne.feyerherm@sorin.com; and if they have transferred the affected products to a third party, to pass this information to the Sorin Group. Further questions may be directed to Sorin Group Customer Service at 1-800-650-2623.
Initiated
2012-09-20
Root cause
Mixed-up of materials/components
Terminated
2013-01-28
Distribution
Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.
Recall number
Z-0167-2013
Recall status
Terminated
CDRH record ID
113636
FDA classified
2012-10-29
Recall event number
63381

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗