Skip to content

Public FDA record

Z-0168-2011: Volcano s5/s5i imaging system Family Software version 3.2, Part Numbers 435-0602.01 to 05, including Chro

Recall record Z-0168-2011 for Volcano Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0168-2011
Company in FDA record
Volcano Corporation
FDA classification date
Nov 1, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software defect results in improper image brightness, which may result in introduction of artifacts, which may cause misclassification of results and affect classification of tissue types. This may result in misdiagnosis and inappropriate treatment.

Source facts

Action
Volcano issued a Voluntary Recall of Volcano s5 Family Software version 3.2 letter dated June 30, 2010, identifying the software defect and actions to be taken by Volcano. Field Service Engineers will visit each site and reinstall the previous software version, which will resolve the issue. Customers can contact Volcano at 916 281-2790.
Initiated
2010-06-29
Root cause
Software design
Terminated
2010-12-22
Distribution
Worldwide Distribution: USA, and the countries of France, Germany, Japan, and the UK.
Recall number
Z-0168-2011
Recall status
Terminated
CDRH record ID
94519
FDA classified
2010-11-01
Recall event number
56794

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗