Software defect results in improper image brightness, which may result in introduction of artifacts, which may cause misclassification of results and affect classification of tissue types. This may result in misdiagnosis and inappropriate treatment.
Source facts
Action
Volcano issued a Voluntary Recall of Volcano s5 Family Software version 3.2 letter dated June 30, 2010, identifying the software defect and actions to be taken by Volcano. Field Service Engineers will visit each site and reinstall the previous software version, which will resolve the issue.
Customers can contact Volcano at 916 281-2790.
Initiated
2010-06-29
Root cause
Software design
Terminated
2010-12-22
Distribution
Worldwide Distribution: USA, and the countries of France, Germany, Japan, and the UK.
Recall number
Z-0168-2011
Recall status
Terminated
CDRH record ID
94519
FDA classified
2010-11-01
Recall event number
56794