Open original FDA record ↗Recall record
- FDA identifier
- Z-0168-2013
- Company in FDA record
- Alphatec Spine, Inc.
- FDA classification date
- Oct 29, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws, therefore rendering such screws nonfunctional.
Source facts
- Action
- Alphatec Spine sent an Urgent Medical Device Recall Notification letter dated September 25, 2012, to all affected customers. The letter identifed the product, the problem, and the action to be taken by the customer. Customers were instructed to contact Alphatec Spine's Customers Service Department (customerservice@alphatecspine.com) immediately for instructions on how to return the affected product. Customers were asked to fill out the lower portion of the letter and fax to (760) 431-0289 to confirm that they have read the notification and will take all necessary removal actions. For questions customers were instructed to call 1-800-922-1356. For questions regarding this recall call 760-431-9286.
- Initiated
- 2012-09-25
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-03-29
- Distribution
- Nationwide Distribution including AZ, IL, CA, SC, TN, TX, GA, CT, NY, FL, OK, LA, MO, NV,and UT.
- Recall number
- Z-0168-2013
- Recall status
- Terminated
- CDRH record ID
- 113646
- FDA classified
- 2012-10-29
- Recall event number
- 63385