Open original FDA record ↗Recall record
- FDA identifier
- Z-0168-2014
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Nov 5, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the sof
Source facts
- Action
- Roche sent an Urgent Medical Device Correction letter dated September 16, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Roche Diagnostics recommends tracking all ISE counters and manually requesting the corresponding Service Actions listed in the notification be monitored until updated software becomes available during Q4 2013. If the affected product was moved to another site, the notification is to be forwarded. consignees were asked to complete the attached faxback form and fax it to 1-877-266-0997 and file the Urgent Medical Device Correction (UMDC) for future reference. Question should be directed to Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-428-2336 For questions regarding thisr ecall call 317-576-3911.
- Initiated
- 2013-09-16
- Root cause
- Software Design Change
- Terminated
- 2015-05-12
- Distribution
- Nationwide Distribution and Puerto Rico
- Recall number
- Z-0168-2014
- Recall status
- Terminated
- CDRH record ID
- 121816
- FDA classified
- 2013-11-05
- Recall event number
- 66239