potential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option.
Source facts
Action
Siemens sent Safety Advisory Notices, dated May 5th and 22nd, 2015 to all affected customers to inform them about an update of the Control Console software to correct defects affecting the Control Console with their digital Linear Accelerator. Customers were instructed to ensure everne at their organization is aware of the information in the Advisory Notice. If product was further distributed, please forward Safety Notice to all those affected.
For questions regarding this recall call 610-219-6300.
Initiated
2015-05-05
Root cause
Software design
Terminated
2017-05-05
Distribution
Nationwide Distribution
Recall number
Z-0168-2016
Recall status
Terminated
CDRH record ID
136546
FDA classified
2015-10-27
Recall event number
71207