Z-0168-2023: Blinq Pediatric Vision Scanner-Pediatric Vision Scanner-intended for automated detection of misalignment

Recall record Z-0168-2023 for Rebiscan, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0168-2023
Company in FDA record
Rebiscan, Inc.
FDA classification date
Nov 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unable to charge the battery pack- battery packs will allow the batteries to be charged if the voltage of any cell is below 2.5V. When this occurs, the battery pack will not accept the charge, could vent and possibly catch fire

Source facts

Action
Rebion issued Urgent Medical Device Removal letter on 10/3/22 via Certified Mail. Letter states reason for recall, health risk and action to take: 1. Return the provided response form acknowledge receipt of this communication. 2. Stop using the device 3. Return the device, along with its power cord and carry case, to Rebion In the event that you have noticed abnormal device behavior related to charging, battery life, or excessive heat while operating the device, then please contact us immediately PLANNED ACTION TO BE TAKEN BY REBION 1. Rebion plans to evaluate and repair each device at no charge. 2. Provide each customer with a postage paid shipping label 3. Upon receipt of the device, Rebion will determine whether the battery pack is affected. 4. Affected battery packs will be replaced. 5. The device will be returned to the customer. FURTHER INFORMATION AND SUPPORT If you need further information or support for this issue, contact your Rebion representative or contact Rebion Customer Service at 877-499-9966.
Initiated
2022-10-03
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of New Zealand, Belgium.
Recall number
Z-0168-2023
Recall status
Open, Classified
CDRH record ID
196320
FDA classified
2022-11-01
Recall event number
90968

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗