Open original FDA record ↗Recall record
- FDA identifier
- Z-0169-2007
- Company in FDA record
- Baxter Healthcare Renal Div
- FDA classification date
- Nov 14, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Meridian pump is less likely to detect small pressure changes, which may indicate the presence of a post-pump tubing kink, when using pre-pump arterial pressure monitoring.
Source facts
- Action
- Urgent Device Correction letters dated 9/6/06 were sent to the Meridian accounts on the same date via UPS second day, to the attention of the Hemodialysis Administrator. The letters informed the accounts of the potential risk associated with a post-pump bloodline kink when using pre-pump arterial pressure monitoring. The accounts were provided with a revised Meridian Operator''s Manual which includes the following warnings: 'The lower venous limit should be set as close to the currently displayed venous pressure as conditions allow.' and 'Reliance on the pressure alarms should not substitute for clinical monitoring of the patient for adverse events.' Any technical questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 3, option 2, choice 2. General questions were directed to The Center for One Baxter at 1-800-422-9837.
- Initiated
- 2006-09-06
- Root cause
- Other
- Terminated
- 2006-12-20
- Distribution
- Nationwide and Hong Kong, Korea and China
- Recall number
- Z-0169-2007
- Recall status
- Terminated
- CDRH record ID
- 48483
- FDA classified
- 2006-11-14
- Recall event number
- 36361