Byrd Screw, Self Drilling, may break during usage. BIOMET has determined that the parts were not correctly processed by the vendor , i.e. that the parts were not hardened in the final processing and that lead to weaker than acceptable material.
Source facts
Action
Biomet plans to send letters to customers on February 19, 2009. Customers are told to discontinue use of the item and return it to Biomet Microfixation. The firm plans to follow up with an e-mail to their customers. Questions pertaining to this issue, should be directed to Christy Cain at 1.800.874.7711 or 904.741.4400.
Initiated
2009-02-19
Root cause
Nonconforming Material/Component
Terminated
2012-05-03
Distribution
AZ, CA, GA, IL, IN, NM, NY, TX, VA, Puerto Rico and Canada.
Recall number
Z-0169-2010
Recall status
Terminated
CDRH record ID
83533
FDA classified
2009-11-10
Recall event number
52639