Skip to content

Public FDA record

Z-0169-2013: MOSAIQ MOSAIQ is an image-enabled electronic medical record system (EMR) used for oncology workflow ma

Recall record Z-0169-2013 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0169-2013
Company in FDA record
Elekta, Inc.
FDA classification date
Oct 30, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

A patient was seriously mistreated after a user made and saved an inadvertent change to the definition of a treatment filed in MOSAIQ.

Source facts

Action
The firm, MOSAIQ, sent an "IMPORTANT SAFETY NOTICE- LINMSQ0003" was sent to consignees/customers on September 18, 2012. The notice described the product, problem and actions to be taken. The customers were instructed to check their Department Setup configuration settings, and re-enable the Field Edit Delta Display feature if they have disabled it. Note: sites can take this action without an upgrade/patch. The customers were also instructed to complete and return the CONFIRMATION OF RECEIPT form via fax to: 702-992-5002 in US or +44 (0)1293 654401 in OUS, Attn: Elekta OIS Support; scan and email to: support@impac.com or europe.support@impac.com; Attn: Elekta OIS Support and/or mail to: C/O Elekta OIS Support, 2310 Corporate Circle, Suite 275, Henderson, NV, USA; or C/O Elekta OIS Support, Elekta Limited, Linac House, Fleming Way, Crawley, West Sussex, RH10 9RR, UK. (Note: Product upgrade/patch is not mandatory or even necessary to reduce the related risk.) For questions, concerns, and requests for upgrade, email: support@impac.com (North America and Rest of World) or europe.support@impac.com (Europe) or call Vice President of RA/QA RNA at 770-670-2548.
Initiated
2012-09-18
Root cause
Device Design
Terminated
2013-06-06
Distribution
Worldwide distribution: USA (nationwide) and country of Canada.
Recall number
Z-0169-2013
Recall status
Terminated
CDRH record ID
113515
FDA classified
2012-10-30
Recall event number
63335

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗