Z-0169-2023: THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based c

Recall record Z-0169-2023 for Johnson & Johnson Surgical Vision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0169-2023
Company in FDA record
Johnson & Johnson Surgical Vision, Inc.
FDA classification date
Nov 2, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.

Source facts

Action
On 09/07/2022, recall notices were mailed and emailed to customers requesting the following: 1) Ensure that anyone in your facility who needs to be aware of this notification reads the notice carefully. 2) If any of the affected products have been forwarded to another facility, contact that facility, and provide a copy of this letter to the relevant personnel. 3) Discard/Destroy the device as per your standard disposal process, ensuring it cannot be used. 4) Complete and return the business response form. Questions can be directed to the recalling firm representative. The following email address was provided: OneMD-Field-Actions@its.jnj.com
Initiated
2022-09-07
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY.
Recall number
Z-0169-2023
Recall status
Open, Classified
CDRH record ID
196121
FDA classified
2022-11-02
Recall event number
90887

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗