Skip to content

Public FDA record

Z-0169-2026: Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Numbe

Recall record Z-0169-2026 for Spacelabs Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0169-2026
Company in FDA record
Spacelabs Healthcare, Inc.
FDA classification date
Oct 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

Source facts

Action
On 09/05/2025, the firm email an "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them that the DVI cables which are used for connecting the Qube (91390) or Xprezzon (91393) to the 94267-19, 94267-L19, 94267-L17 and 94267-L15 displays, do not meet the EMI/EMC electrical safety standards specifically for electromagnetic shielding and may pose serious risks to patient safety by interfering with the operation of another device, particularly in environments where electrically sensitive medical devices are in use. Customers are instructed to acknowledge receipt and understanding of this notification and request replacement cables as soon as possible, however no later than 31-Dec-2025. Do not replace cables while a patient is admitted to the monitor to mitigate any potential disruption in active patient monitoring. Any affected cables not currently in use should not be placed into service. For questions and/or assistance with this recall, contact: Spacelabs Technical Support Spacelabs Healthcare, Inc 35301 SE Center Street Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support Email: Support@spacelabs.com
Initiated
2025-09-05
Root cause
Device Design
Distribution
Worldwide - United States Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of AFG, ARE, ARG, AUS, AUT, BGD, BHR, BOL, BRA, BRN, CAN, CHE, CHL, CHN, COL, CRI, CYP, CZE, DOM, DZA, ECU, EGY, ESP, FRA, GBR, GRC, GTM, GUM, HKG, IDN, IND, IRL, IRN, ISR, ITA, JOR, KWT, LBN, MAR, MEX, MYS, NLD, NZL, OMN, PAK, PAN, PER, PHL, PLS, POL, PRI, PRT, PRY, QAT, ROM, SAU, SGP, SLV, SRB, THA, TTO, TUN, TUR, TWN, VNM.
Recall number
Z-0169-2026
Recall status
Open, Classified
CDRH record ID
216221
FDA classified
2025-10-16
Recall event number
97616

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗