Open original FDA record ↗Recall record
- FDA identifier
- Z-0170-2010
- Company in FDA record
- Advanced Neuromodulation Systems, Inc
- FDA classification date
- Nov 10, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Product's external charging system on occasion was failing to communicate with the Implantable Pulse Generator preventing re-charging of the pain relief system.
Source facts
- Action
- 12/22/2006, Marketing Update 5015 was issued to field sales representatives providing information on improving communication between the Internal Pulse Generator and its charger. May 15, 2007, Marketing Update 5022 was issued to field sales representatives to reiterate avoiding contact between the charger antenna and conductive materials and place the preferred side of the charging antenna face-down over the implant. Troubleshooting Guide was created in January 2008 for field sales representatives and technical services departments regarding isolation of the communication issue to the external devices instead of the implanted component. March 25, 2008, the Clinician's Manual was updated adding a recommendation for a minimum depth of pocket for device implantation.
- Initiated
- 2006-12-22
- Root cause
- Device Design
- Terminated
- 2010-07-23
- Distribution
- Worldwide
- Recall number
- Z-0170-2010
- Recall status
- Terminated
- CDRH record ID
- 84736
- FDA classified
- 2009-11-10
- Recall event number
- 53128