Z-0170-2024: Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708),

Recall record Z-0170-2024 for Asensus Surgical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0170-2024
Company in FDA record
Asensus Surgical Inc
FDA classification date
Oct 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.

Source facts

Action
Asensus Surgical issued an URGENT MEDICAL DEVICE RECALL to its consignees on 09/15/2023 by mail. The notice explained the problem, risk to health, and requested users to cease use of the device until software version 2.7.5 is installed.
Initiated
2023-09-15
Root cause
Software design
Terminated
2024-09-26
Distribution
US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania
Recall number
Z-0170-2024
Recall status
Terminated
CDRH record ID
203513
FDA classified
2023-10-30
Recall event number
93129

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗