Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.
Source facts
Action
Asensus Surgical issued an URGENT MEDICAL DEVICE RECALL to its consignees on 09/15/2023 by mail. The notice explained the problem, risk to health, and requested users to cease use of the device until software version 2.7.5 is installed.
Initiated
2023-09-15
Root cause
Software design
Terminated
2024-09-26
Distribution
US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania
Recall number
Z-0170-2024
Recall status
Terminated
CDRH record ID
203513
FDA classified
2023-10-30
Recall event number
93129