Z-0171-2018: BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD

Recall record Z-0171-2018 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0171-2018
Company in FDA record
Becton Dickinson & Co.
FDA record creation date
Nov 25, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. As the product should not be used when quality control (QC) fails, it is unlikely that there would be any impact on patient results, however, if a customer does not perform QC testing and uses the panel, there is a potential to incorrectly treat a Staphylococcus infection with clindamycin based on the iMLSb test result. This issue can be detected 100% of the time when used per package insert instructions with iMLSb positive strain S. aureus BAA-977.

Source facts

Action
URGENT PRODUCT RECALL LETTER (dated May 2017) were sent to customers via a UPS mailing on May 11, 2017. Customers were instructed to discontinue use of the affected lot and discard any remaining product. BD will replace the discarded inventory. Please complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. Please complete and fax the form to: Attention: Regulatory Compliance, Fax: 410-316-4258 or e-mail to: BDRC2@bd.com. For further assistance, please contact BD (Customer/Technical) Support at 1-800-638-8663 in the USA.
Initiated
2017-05-11
Root cause
Manufacturing material removal
Terminated
2018-02-06
Distribution
CT, VA, PA, TX, FL, NC, WI, NM
Recall number
Z-0171-2018
Recall status
Terminated
CDRH record ID
155855
FDA record created
2017-11-25
Recall event number
77370

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗