Z-0171-2020: Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.

Recall record Z-0171-2020 for ICU Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0171-2020
Company in FDA record
ICU Medical Inc
FDA classification date
Oct 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.

Source facts

Action
On 8/1/2018 a "Urgent Medical Device Correction" letter was mailed to consignees via traceable mailing. The medical device correction letter informed consignees of the following: 1. To reduce the likelihood of the issue occurring at the bedside, you can consider not utilizing the option to "Force Pump to Accept Drug Library During Power Down" when finalizing the CDL with MedNet. In doing so, when powering down the infuser the clinician will have the option to install the CDL or not, based on the clinical situation. 2. ICU Medical will release a software update to correct the issue and will contact you when the update is available. 3. . Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular 4.Please inform potential users in your organization of this notification, complete the attached response form and return it to the fax number (1-888-628-6849 or e-mail address (ICUMedical3618@stericycle.com) on the form, even if you do not have the affected product. 5. For further inquiries, please contact ICU Medical using the information provided below: Global Complaint Management - 1-844-654-7780 or productcomplaintsPP@icumed.com Technical Support Center - 1-800-241-4002, option 4 or TSC.support@icumed.com
Initiated
2018-08-01
Root cause
Software change control
Terminated
2021-04-27
Distribution
US nationwide distribution, and Canada
Recall number
Z-0171-2020
Recall status
Terminated
CDRH record ID
175374
FDA classified
2019-10-22
Recall event number
83480

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗