Z-0172-04: ARCHITECT Processing Module, manufactured by Abbott Laboratories, Irving, Texas

Recall record Z-0172-04 for Abbott Laboratories, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0172-04
Company in FDA record
Abbott Laboratories, Inc
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

If the interlock cable connected to the door lid sensor is disconnected or fails, the lid is detected as closed even though it may be open.

Source facts

Action
Firm sent technical service bulletins with corrective instructions to consignees with all existing ARCHITECT i 2000 processing modules in February 2003.
Initiated
2002-05-08
Root cause
Other
Terminated
2003-11-26
Distribution
States of Virginia, Illinois, New York, Texas, Oklahoma, Pennsylvania, ILlinois, Michigan, California, New Hampshire, Ohio, Florida, Maryland, Nevada, Georgia, Minnesota, Colorado, Arizona, Missouri, Louisiana, Oregon, Wisconsin, Hawaii, Connecticut, New Hampshire, Massachusetts, Washington, South Carolina, Mississippi, Alabama, and Tennessee, and the following foreign countries: Austria, Denmark, Germany, Italy, Poland, South Africa, Sweden, England, Colombia, Australia, Japan, New Zealand, Belgium, Finland, Greece, Netherlands, Portugal, Romania, Turkey, Canada, Mexico, Hong Kong, Korea, Taiwan, Czech Republic, France, Israel, Norway, Saudi Arabia, Spain, United Arab Emirates, Brazil, Venezuela, India, Malaysia, and Thailand.
Recall number
Z-0172-04
Recall status
Terminated
CDRH record ID
29383
FDA classified
2004-07-20
Recall event number
27180

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗