Z-0172-06: Advanced D-Dimer used with the Sysmex CA-1500, CA-7000, CA-500 Systems when using the AKIMA method.

Recall record Z-0172-06 for Dade Behring , Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0172-06
Company in FDA record
Dade Behring , Inc.
FDA classification date
Nov 18, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Reports of false negative results on Sysmex Coagulation System.

Source facts

Action
On 7/29/05, the recalling firm issued a recall letter to their accounts informing them of the probelm. The letter also provided instruction on how to prevent the probelm from occurring - change from the AKIMA method to the Point ot Point method.
Initiated
2005-07-29
Root cause
Other
Terminated
2006-03-01
Distribution
The product was shipped to medical facilities nationwide.
Recall number
Z-0172-06
Recall status
Terminated
CDRH record ID
41029
FDA classified
2005-11-18
Recall event number
32887

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗