Z-0172-2018: BRAINLAB EXACTRAC VERO, Model/Catalog Numbers: 46228 EXACTRAC VERO 3.5 46238 EXACTRAC VERO 3.5 46216 EX

Recall record Z-0172-2018 for Brainlab AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0172-2018
Company in FDA record
Brainlab AG
FDA record creation date
Nov 28, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This may result in the treatment couch not moving to the exact planned position resulting in a deviation between the planned and the treatment target position. This deviation may not be clearly visible to the user if no verification image of the patient position is acquired after this couch shift has been performed.

Source facts

Action
The firm notified their consignees by email on November 3, 2017. "Instructions to customers what to do with the recalled product. For instructions for the user how to address this issue, and to avoid a patient and treatment target position potentially different than intended due to this issue, please also refer to the User Corrective Action section in the attached Field Safety Notice / Product Notification letter CAPA-20171011-001985: 1. After applying a couch shift always perform a verification image of the patient position before enabling the treatment beam. 2. Do not use the Virtual Isocenter function if a verification of the patient position in the treatment isocenter is not feasible. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning in the Clinical User Guide ExacTrac Vero: Brainlab highly recommends performing an X-ray verification after each change to the couch/Robotics Position." For further questions, please call 1 (800) 784-7700.
Initiated
2017-11-03
Root cause
Software design
Terminated
2020-12-14
Distribution
Worldwide Distribution - US Distribution and to the countries of : Belgium, Canada, France, Germany, Italy, Japan and S. Korea
Recall number
Z-0172-2018
Recall status
Terminated
CDRH record ID
159653
FDA record created
2017-11-28
Recall event number
78488

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗