Open original FDA record ↗Recall record
- FDA identifier
- Z-0172-2018
- Company in FDA record
- Brainlab AG
- FDA record creation date
- Nov 28, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This may result in the treatment couch not moving to the exact planned position resulting in a deviation between the planned and the treatment target position. This deviation may not be clearly visible to the user if no verification image of the patient position is acquired after this couch shift has been performed.
Source facts
- Action
- The firm notified their consignees by email on November 3, 2017. "Instructions to customers what to do with the recalled product. For instructions for the user how to address this issue, and to avoid a patient and treatment target position potentially different than intended due to this issue, please also refer to the User Corrective Action section in the attached Field Safety Notice / Product Notification letter CAPA-20171011-001985: 1. After applying a couch shift always perform a verification image of the patient position before enabling the treatment beam. 2. Do not use the Virtual Isocenter function if a verification of the patient position in the treatment isocenter is not feasible. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning in the Clinical User Guide ExacTrac Vero: Brainlab highly recommends performing an X-ray verification after each change to the couch/Robotics Position." For further questions, please call 1 (800) 784-7700.
- Initiated
- 2017-11-03
- Root cause
- Software design
- Terminated
- 2020-12-14
- Distribution
- Worldwide Distribution - US Distribution and to the countries of : Belgium, Canada, France, Germany, Italy, Japan and S. Korea
- Recall number
- Z-0172-2018
- Recall status
- Terminated
- CDRH record ID
- 159653
- FDA record created
- 2017-11-28
- Recall event number
- 78488