The Tina System was released to the customer with OOS values for the Pressure Displays Accuracy and Leakage Rate Test.
Source facts
Action
Baxter Healthcare Corporation sent a service technician to the lone customer site on September 19, 2011 to correct the affected product. The technician executed all tests as established in the Service Manual, with acceptable results.
Initiated
2011-09-19
Root cause
Employee error
Terminated
2011-11-07
Distribution
Distributed in Puerto Rico only.
Recall number
Z-0173-2012
Recall status
Terminated
CDRH record ID
104698
FDA classified
2011-11-07
Recall event number
60183