Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Oct 22, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Medtronic is recalling three lots of unused C315HIS Delivery Catheters due to reports of an inability to pass a lead through the catheter.
Source facts
Action
Written notification letter titled, "MEDICAL DEVICE RECALL C315HIS DELIVERY CATHETER" was sent to all consignees on 09/09/2021 via 2-day UPS delivery. The letter instructs the consignee to take the following actions: "Check your inventory for the catheters identified in Table 1 above. If the product has been consumed, please complete the enclosed Customer Confirmation Form and email to xxxxx@medtronic.com. No further actions are required.
" If there is unused identified product in your inventory, quarantine and return it to Medtronic. Contact Medtronic Customer Service at xxxxxxx to initiate a product return. Your local Medtronic Representative can assist you in the return of this product.
o Note: Root cause is under investigation for this issue. Return of the identified product will aid the investigation.
Please complete the enclosed Customer Confirmation Form and email to xxxxxx@medtronic.com." The letter also instructs the consignee to share the notice with all those who need to be made aware with their organization or to any organization where the potentially affected products have been transferred.
Initiated
2021-09-09
Root cause
Under Investigation by firm
Distribution
Distributed internationally to Australia, Brunei Darussalam, China, Germany, Hong Kong, Indonesia, Malaysia, Netherlands, Singapore, and Taiwan.
Recall number
Z-0173-2022
Recall status
Open, Classified
CDRH record ID
189295
FDA record created
2021-10-22
Recall event number
88688