Skip to content

Public FDA record

Z-0173-2023: BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, B

Recall record Z-0173-2023 for Becton, Dickinson and Company, BD Biosciences. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0173-2023
Company in FDA record
Becton, Dickinson and Company, BD Biosciences
FDA classification date
Nov 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant.

Source facts

Action
On 10/03/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via FedEx tracked letters and emails informing customers, through internal investigations, the labels applied to the BD Trucount tubes are becoming detached due to an adhesive change made at a supplier. Customers are instructed to: 1. Immediately review their inventory for the specific catalog and lot numbers. 2. For any tubes where label detachment is observed, BD recommends that the clinician remove the label from the tube and, using a permanent marker, record the tube lot number as well as any other relevant information directly on the tube. Removal of the label prevents it from interfering with any auto-loading sample preparation and analysis systems, as well as from sticking to other tubes. For questions, contact North American Regional Complaint Center at: Phone: 1-844-8BD-LIFE (1-844-823-5433), Say "Recall" when prompted Mon-Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
Initiated
2022-10-03
Root cause
Under Investigation by firm
Distribution
U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay
Recall number
Z-0173-2023
Recall status
Open, Classified
CDRH record ID
196097
FDA classified
2022-11-03
Recall event number
90866

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗