Open original FDA record ↗Recall record
- FDA identifier
- Z-0174-04
- Company in FDA record
- Invacare Corporation
- FDA classification date
- Jul 20, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Pull tube on bed may bend or separate causing inoperability of bed or head section to fall.
Source facts
- Action
- The firm began telephoning direct consignees on 6/9/03 and also began sending recall letters to these consignees on 6/9/03. The letter provides a list of affected units shipped to the consignee and the serial number of the affected foot section on the bed. If the consignee indicated they wanted to perform the correction themselves then the indicated number of corrective bracket installation kits are being sent. Also they are given the option of having an Invacare authorized service center perform the correction. This is for units in their possession and also at patient homes. As of October 2003 enough corrective kits have been forwarded to dealers to correct all involved units.
- Initiated
- 2003-06-09
- Root cause
- Other
- Terminated
- 2014-05-08
- Distribution
- involved products were distributed to approximately 1100 dealers in the United Stated and to 54 Canadian dealer accounts.
- Recall number
- Z-0174-04
- Recall status
- Terminated
- CDRH record ID
- 27751
- FDA classified
- 2004-07-20
- Recall event number
- 26479