Open original FDA record ↗Recall record
- FDA identifier
- Z-0174-2010
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Nov 10, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip file on the NAS or HSM will contain fewer images-those which have been archived until the time of the error/interruption.
Source facts
- Action
- The recalling firm issued a "Customer Safety Advisory Notice" dated 9/14/09 to all affected customers via Update Instructions IM101/09/S. The notice informs the customer of the potential issue and provides instructions to avoid its occurrence. Questions regarding this issue are directed to the local SIEMENS Uptime Service Center.
- Initiated
- 2009-09-30
- Root cause
- Software design
- Terminated
- 2012-07-05
- Distribution
- The product was shipped to medical facilities in NY and OH
- Recall number
- Z-0174-2010
- Recall status
- Terminated
- CDRH record ID
- 85877
- FDA classified
- 2009-11-10
- Recall event number
- 53564