Open original FDA record ↗Recall record
- FDA identifier
- Z-0174-2014
- Company in FDA record
- St. Jude Medical
- FDA classification date
- Nov 6, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
St. Jude Medical Cardiovascular and Ablation Technologies Division is performing a voluntary notification regarding selected batches of Coolpath Duo Ablation Catheter MediGuide Enabled products and Safire Duo Ablation Catheter MediGuide Enabled. These units were inadvertently shipped with the incorrect version of the Instructions for Use (IFU).
Source facts
- Action
- The firm, St. Jude Medical (SJM), sent a "Urgent Medical Device Notice" dated 11 October 2013 to its consignees. The letter described the problem, the product involved in the recall and the actions to be taken. The consignees were instructed to discard the IFU received with the batches and replace it with the IFU received in conjunction with the notification letter. SJM has determined that it is acceptable to continue using these devices as long as the product is used in accordance with the IFU. If you have any questions, contact Vice President, Quality, at 651-756-5402.
- Initiated
- 2013-10-11
- Root cause
- Error in labeling
- Terminated
- 2016-03-10
- Distribution
- US distribution: KS, MA, UT,TX, KY, OH.
- Recall number
- Z-0174-2014
- Recall status
- Terminated
- CDRH record ID
- 122676
- FDA classified
- 2013-11-06
- Recall event number
- 66597