Z-0174-2022: Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic Brachyt

Recall record Z-0174-2022 for Xstrahl Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0174-2022
Company in FDA record
Xstrahl Limited
FDA record creation date
Oct 26, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible mis-treatment.

Source facts

Action
The recalling firm issued an Urgent Field Safety Notice dated 4/26/2021 via email on 4/26/2021. The notice explained the issue, impact of the issue, and provided actions to be taken for Concerto 2.x 2 Beam Plans Only, although all customers with the affected software were issued the safety notice. The consignee was informed that until an update to their affected software version is installed, they should follow the actions listed in the letter.
Initiated
2021-04-26
Root cause
Software design
Terminated
2024-01-26
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, KY, NV, OR, PA, TN, and TX. The countries of Austria, Germany, and Switzerland.
Recall number
Z-0174-2022
Recall status
Terminated
CDRH record ID
189414
FDA record created
2021-10-26
Recall event number
88732

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗