Open original FDA record ↗Recall record
- FDA identifier
- Z-0174-2022
- Company in FDA record
- Xstrahl Limited
- FDA record creation date
- Oct 26, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible mis-treatment.
Source facts
- Action
- The recalling firm issued an Urgent Field Safety Notice dated 4/26/2021 via email on 4/26/2021. The notice explained the issue, impact of the issue, and provided actions to be taken for Concerto 2.x 2 Beam Plans Only, although all customers with the affected software were issued the safety notice. The consignee was informed that until an update to their affected software version is installed, they should follow the actions listed in the letter.
- Initiated
- 2021-04-26
- Root cause
- Software design
- Terminated
- 2024-01-26
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, KY, NV, OR, PA, TN, and TX. The countries of Austria, Germany, and Switzerland.
- Recall number
- Z-0174-2022
- Recall status
- Terminated
- CDRH record ID
- 189414
- FDA record created
- 2021-10-26
- Recall event number
- 88732