Z-0174-2023: BD Nexiva Closed IV Catheter System - Single Port REF 383516 20 GA X1.0o in (1.1 x 25 mm) 61 mL/min (3660

Recall record Z-0174-2023 for Becton Dickinson Infusion Therapy Systems Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0174-2023
Company in FDA record
Becton Dickinson Infusion Therapy Systems Inc.
FDA classification date
Nov 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During needle withdrawal, the tip shield may prematurely separate from the catheter adapter and preventing the needle tip from being in the safely shielded (leaving the needle tip exposed).

Source facts

Action
On 10/12/2022, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via FedEx and email, to customers informing them that Becton, Dickinson (BD) confirmed that during withdrawal of the Nexiva product, the tip shield may prematurely separate from the catheter adapter and prevent the needle tip from being safely shielded. As a result, the clinicians could potentially experience a contaminated needlestick injury after withdrawal of the needle. Customers are instructed to: 1. Immediately review their inventory for the specific Catalog (Ref) and Lot number listed above. Destroy all unused product subject to the recall following their institution s process for destruction. 2. For indwelling catheters, no further action is needed. As indicated, these catheters can remain in place and continue to be used per the Instructions for Use. 3. If the affected product was previously used in patient care, no further action is needed. 4. Share this recall notification with all users within their facility network of the product to ensure they are also aware of this recall. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not they have any of the impacted material so that BD may acknowledge their receipt of this notification per FDA requirements and subsequently process your product replacement. For questions - contact North America Regional Complaint Center at Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
Initiated
2022-10-12
Root cause
Storage
Distribution
U.S.: New York, Florida, Maryland, North Carolina, Virginia, Georgia, Rhode Island, New Jersey, Pennsylvania, Illinois, South Carolina, Maine, and Massachusetts. O.U.S.: N/A
Recall number
Z-0174-2023
Recall status
Open, Classified
CDRH record ID
196361
FDA classified
2022-11-03
Recall event number
90972

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗