Open original FDA record ↗Recall record
- FDA identifier
- Z-0174-2024
- Company in FDA record
- American Contract Systems, Inc.
- FDA classification date
- Oct 26, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
Source facts
- Action
- Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 09/01/2023 by email. The notice explained the issue, risk, and requested the user discontinue the use of the I.V. Cath and Adhesive Dermabond and affix a warning label to each kit and case of affected product in inventory. The firm sent an updated letter on 09/22/2023 via email which corrected the product distribution date.
- Initiated
- 2023-09-01
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of FL, IA.
- Recall number
- Z-0174-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203412
- FDA classified
- 2023-10-26
- Recall event number
- 93086