Z-0175-05: Stryker brand Neptune waste management system, Neptune Rover, 15 amp plug; model 0700-003-000.

Recall record Z-0175-05 for Stryker Instruments Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0175-05
Company in FDA record
Stryker Instruments Div. of Stryker Corporation
FDA classification date
Nov 4, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The grounding pin may break inside of the electrical plug, resulting in a shock hazard.

Source facts

Action
Hospital customers will be sent a recall notice dated 11/09/04, new plugs for installation on their units, and instructions for installation.
Initiated
2004-10-12
Root cause
Other
Terminated
2005-03-28
Distribution
United States and Canada.
Recall number
Z-0175-05
Recall status
Terminated
CDRH record ID
35627
FDA classified
2004-11-04
Recall event number
30208

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗