Open original FDA record ↗Recall record
- FDA identifier
- Z-0175-2007
- Company in FDA record
- Asahi Medical Co., Ltd
- FDA classification date
- Nov 14, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle.
Source facts
- Action
- Asahi sent Safety Alert letters on 10/12/06 to their re-use dialyzer customers advising them of the reports of difficulty separating the blood set from the dialyzer and not to use the Gambro Cartridge Blood Set with Asahi Reuse/Multiple Use Dialyzers in a reuse situation. Enclosed were copies of the temporary revision Instructions For Use with the following warning added at the top of the page: 'Additional Warning for Reuse/Multiple Use Dialyzers: Ashai Reuse/Multiple use dialyzers must not be used in connection with the Gambro cartridge Blood Set in a reuse environment.' Any questions were directed to Karen Borling, RN at 847-498-8500.
- Initiated
- 2006-10-12
- Root cause
- Other
- Terminated
- 2007-11-28
- Distribution
- Nationwide
- Recall number
- Z-0175-2007
- Recall status
- Terminated
- CDRH record ID
- 48646
- FDA classified
- 2006-11-14
- Recall event number
- 36430