Open original FDA record ↗Recall record
- FDA identifier
- Z-0175-2008
- Company in FDA record
- Electric Mobility Corp
- FDA classification date
- Nov 21, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Seat and/or lift actuator problems: 1) Models equipped with a Seat Lift, the mast screw and washer holding the Seat Lift Actuator in place will loosen break off or fall out; 2) Models assembled with Seat Lift Actuator or Seat Post Mounting Bolts, the seat may wobble or the seat will break off from the vehicle; 3) Models using Seat-On Lifting brackets with a hoist to raise an EMC vehicle equipped w
Source facts
- Action
- Consumers were sent an Urgent Medical Device Recall Notification on October 12, 2007, requesting that they immediately respond to the Medical Device Recall by calling EMC at 1-800-814-4365, or by returning an enclosed Recall Response Form, or by responding on-line at www.EMCIT.NET/801B. The Notification address three important safety issues on 1) Seat Lift Actuator Mast Screws and Washers; 2) Seat Lift Actuator or Seat Post Mounting Bolts and; 3) Incorrect Seat-On Lift Bracket Installation. Corrections to the devices will be made by the nearest Service Representatives, by scheduled in-home repairs, or return of the rear section of the scooter with EMC providing shipping material and freight payment.
- Initiated
- 2007-03-29
- Root cause
- Process change control
- Terminated
- 2014-04-08
- Distribution
- Nationwide: to consumers via direct purchase from Electric Mobility Corporation.
- Recall number
- Z-0175-2008
- Recall status
- Terminated
- CDRH record ID
- 52042
- FDA classified
- 2007-11-21
- Recall event number
- 37880