Z-0175-2020: ApexPro Telemetry Server System. Also identified as Modification To: ApexPro Telemetry System and ApexPr

Recall record Z-0175-2020 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0175-2020
Company in FDA record
GE Healthcare, LLC
FDA classification date
Oct 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

May not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions.

Source facts

Action
GE Healthcare notified customers on about 08/30/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were made aware of the below six (6) potential safety issues along with corresponding safety instructions. Safety Issue #1: Telemetry patients monitored on a Central Station or in combination monitoring mode with a CARESCAPE B850/650/450, DASH 3000/4000/5000, or SOLAR 8000M/I bedside monitor under certain conditions can experience the following: ECG arrhythmias configured to MEDIUM (WARNING) or lower alarm priority occurring before and after a NO TELEM condition may not re-activate after the NO TELEM condition resolves. This issue may result in delayed treatment for a potentially life-threatening ECG arrhythmia event. There have been no reported injuries as a result of this issue. Safety instructions included that the monitor can continue being used under these instructions: 1. Configure the VTACH arrhythmia to HIGH (CRISIS) alarm priority. 2. Any ECG arrhythmia of clinical concern should be configured to HIGH (CRISIS) alarm priority. 3. Make sure that the ECG HR alarm is set to priority of MEDIUM (WARNING) or higher to ensure proper alarming. 4. Ensure that your ECG HR limits are appropriately configured for your patient population. Safety Issue #2: Telemetry patients monitored on a Central Station or in combination monitoring mode with a CARESCAPE B850/650/450, DASH 3000/4000/5000 or SOLAR 8000M/I bedside monitor under certain conditions can experience the following: A NO TELEM INFORMATIONAL (MESSAGE) indication may remain displayed on the Central Station instead of a HIGH (CRISIS) LEADS FAIL or a MEDIUM (WARNING) NO TELEM alarm. This issue can result in delayed treatment for a potentially a life-threatening event, as a failure to notice prolonged loss of ECG monitoring. There have been no reported injuries as a result of this issue. Safety instructions included that the monitor can continue being used by following these instructions: For a N
Initiated
2019-08-30
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. The countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BERMUDA, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUAM, GUATEMALA, HONDURAS, HONG KONG, HUNGARY, INDIA, Indonesia, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MACEDONIA, Malaysia, MEXICO, MOROCCO, Myanmar, NETHERLANDS, New Zealand, NORWAY, PANAMA, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, QATAR, ROMANIA, Russia, SAUDI ARABIA, Singapore, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TAJIKISTAN, Thailand, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, Viet Nam, YUGOSLAVIA.
Recall number
Z-0175-2020
Recall status
Open, Classified
CDRH record ID
176411
FDA classified
2019-10-23
Recall event number
83813

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗