Z-0175-2022: Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD se

Recall record Z-0175-2022 for AB SCIEX. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0175-2022
Company in FDA record
AB SCIEX
FDA record creation date
Oct 26, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making incorrect conclusions using incorrect results.

Source facts

Action
On October 4, 2021, the firm distributed Urgent Medical Device Correction letters to affected consignees. Customers were provided with temporary actions that could be implemented for safe handling of the recalled software to prevent the reporting of incorrect results to healthcare providers resulting in incorrect conclusions: 1. Avoid setting the IS (Internal Standard) column as hidden AND leaving it empty 2. Review values in the IS Concentration column 3. Follow lab SOPs to ensure quality control SCIEX is working on the hotfix for Cliquid MD Software version 3.4 and will contact affected customers as soon as it is ready The hotfixes may be downloaded free of charge directly from SCIEX s website, at the following location: https://sciex.com/software-support/software-downloads Software release notes will provide detailed installation instructions for installing the Hotfix. If you have any questions, please contact customer service at 1-877-740-2129, option 1, and then option 4, or your local SCIEX support number. You may also submit questions through their web portal at https://sciex.com/request-support.
Initiated
2021-10-04
Root cause
Software Design Change
Terminated
2024-01-23
Distribution
Worldwide distribution - US Nationwide distribution in the states of District of Columbia, Florida, Illinois, Louisiana, Maryland, Michigan, Missouri, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Tennessee, Utah, Wisconsin and the countries of Austria, Finland, Germany, Hong Kong, Italy, Japan, Norway, Spain, Switzerland, United Kingdom, Italy, France, and Japan.
Recall number
Z-0175-2022
Recall status
Terminated
CDRH record ID
189437
FDA record created
2021-10-26
Recall event number
88748

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗