The manufacturer's instructions for using radiation treatment software may cause healthcare practitioner to mistreat the patient.
Source facts
Action
The recalling firm notified consignees by Customer Technical Bulletin on 5/21/03 and additionally by 'URGENT MEDICAL DEVICE CORRECTION LETTER' on 10/31/03. The notifications caution consignees about the need to fully understand scaling functionality and to check scaling for every treatment plan. The notification further explains that the next software release will be modified to include a screen message that appears whenever automatic magnification appears.
Initiated
2003-10-31
Root cause
Other
Terminated
2003-12-15
Distribution
Software was distributed to end users nationwide and internationally.
Recall number
Z-0176-03
Recall status
Terminated
CDRH record ID
30210
FDA classified
2003-12-03
Recall event number
27629