Skip to content

Public FDA record

Z-0176-05: Alto implantable cardioverter defibrillator

Recall record Z-0176-05 for ela Medical Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0176-05
Company in FDA record
ela Medical Llc
FDA classification date
Nov 4, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Alto Implantable Defibrillators Models 614 and 615. These models can experience sudden no output, and early end of life characteristics due to metal migration in their controlling hybrids, causing high current drain leading to premature battery depletion.

Source facts

Action
The company observed occurrences of premature battery depletion and/or prolonged charge time for some Alto 614 and Alto 615 devices. The firm is requesting physician's assistance to monitor the performance of a limited number of implanted Alto 614 and 615 devices.
Initiated
2004-10-21
Root cause
Other
Terminated
2005-11-14
Distribution
Nationwide. Devices were received from the manufacturer in France. After received in the US, No devices were distributed from the US to foreign consignees. 95 physicians received the recalled product. ELA Medical has contacted physicians who follow patients implanted with the devices that are the subject of the correction. It should be noted that in many cases, the physician following the patient is not the physician who implanted the device. Because ELA is requesting assistance to monitor the performance of implanted devices, this notification is directed at the physicians who are following the patients.
Recall number
Z-0176-05
Recall status
Terminated
CDRH record ID
35594
FDA classified
2004-11-04
Recall event number
30273

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗