Open original FDA record ↗Recall record
- FDA identifier
- Z-0176-05
- Company in FDA record
- ela Medical Llc
- FDA classification date
- Nov 4, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Alto Implantable Defibrillators Models 614 and 615. These models can experience sudden no output, and early end of life characteristics due to metal migration in their controlling hybrids, causing high current drain leading to premature battery depletion.
Source facts
- Action
- The company observed occurrences of premature battery depletion and/or prolonged charge time for some Alto 614 and Alto 615 devices. The firm is requesting physician's assistance to monitor the performance of a limited number of implanted Alto 614 and 615 devices.
- Initiated
- 2004-10-21
- Root cause
- Other
- Terminated
- 2005-11-14
- Distribution
- Nationwide. Devices were received from the manufacturer in France. After received in the US, No devices were distributed from the US to foreign consignees. 95 physicians received the recalled product. ELA Medical has contacted physicians who follow patients implanted with the devices that are the subject of the correction. It should be noted that in many cases, the physician following the patient is not the physician who implanted the device. Because ELA is requesting assistance to monitor the performance of implanted devices, this notification is directed at the physicians who are following the patients.
- Recall number
- Z-0176-05
- Recall status
- Terminated
- CDRH record ID
- 35594
- FDA classified
- 2004-11-04
- Recall event number
- 30273