Reason stated in the source
Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data.
Source facts
- Action
- Datascope sent out a Field Correction notice on 9/7/2005 to all Hospital administrators who have the monitors. This was sent out via certified mail, return receipt requested. A Datascope Service Rep will conduct and document the corrective action during a visit to the customer.
- Initiated
- 2005-09-07
- Root cause
- Other
- Terminated
- 2007-06-06
- Distribution
- The monitors have been distributed domestically and internationally to hospitals and clinics. There are 10 Passport 2 monitors distributed to Naval Hospital, Camp Pendleton, CA. There are 12 Spectrum monitors distributed to two US Govt facilities: VA Medical Center, Hines, IL and VA Medical Center, Syracuse, NY.
- Recall number
- Z-0176-06
- Recall status
- Terminated
- CDRH record ID
- 42350
- FDA classified
- 2005-11-23
- Recall event number
- 33724