Skip to content

Public FDA record

Z-0176-2010: RESPIRONICS V60 VENTILATOR

Recall record Z-0176-2010 for Respironics California Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0176-2010
Company in FDA record
Respironics California Inc
FDA classification date
Nov 11, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated to correct an issue with a component on the V60 Ventilator Power Management (PM) Printed Circuit Board (PCB). A component on the PCB has the potential to break off or become damaged if the PCB is removed from the ventilator. If this occurs, the ventilator may not properly switch between back up battery power and AC power. This can result in power loss during ventilation fo

Source facts

Action
The recall notification was initiated October 26, 2009 with the firm mailing a Urgent Medical Device Field Correction letter to the affected customers. The letter informed the customers of the reason for the notice and how Respironics Field Service Technicians were going to correct this issue. The letter also gave the customers instructions to follow pending the completion of the field inspection/replacement. For questions or additional information contact Respironics US Customer Service at 1.800.345.6443, Option #5 then Option #1.
Initiated
2009-10-26
Root cause
Device Design
Terminated
2011-10-13
Distribution
Nationwide and Internationally to : Libya, Hong Kong, France, Italy, India, Spain, United Kingdom, Germany, Netherlands and Saudi Arabia
Recall number
Z-0176-2010
Recall status
Terminated
CDRH record ID
86309
FDA classified
2009-11-11
Recall event number
53676

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗