Open original FDA record ↗Recall record
- FDA identifier
- Z-0176-2010
- Company in FDA record
- Respironics California Inc
- FDA classification date
- Nov 11, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated to correct an issue with a component on the V60 Ventilator Power Management (PM) Printed Circuit Board (PCB). A component on the PCB has the potential to break off or become damaged if the PCB is removed from the ventilator. If this occurs, the ventilator may not properly switch between back up battery power and AC power. This can result in power loss during ventilation fo
Source facts
- Action
- The recall notification was initiated October 26, 2009 with the firm mailing a Urgent Medical Device Field Correction letter to the affected customers. The letter informed the customers of the reason for the notice and how Respironics Field Service Technicians were going to correct this issue. The letter also gave the customers instructions to follow pending the completion of the field inspection/replacement. For questions or additional information contact Respironics US Customer Service at 1.800.345.6443, Option #5 then Option #1.
- Initiated
- 2009-10-26
- Root cause
- Device Design
- Terminated
- 2011-10-13
- Distribution
- Nationwide and Internationally to : Libya, Hong Kong, France, Italy, India, Spain, United Kingdom, Germany, Netherlands and Saudi Arabia
- Recall number
- Z-0176-2010
- Recall status
- Terminated
- CDRH record ID
- 86309
- FDA classified
- 2009-11-11
- Recall event number
- 53676