Open original FDA record ↗Recall record
- FDA identifier
- Z-0176-2020
- Company in FDA record
- Brainlab AG
- FDA record creation date
- Oct 23, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Brainlab Patient Data Manager software (PDM) in combination with the Brainlab iPlan CMF/Cranial/ENT/Spine software 2.1.1, 2.2.0, 2.2.1, 2.3.0, 2.3.1, 2.4.0, 2.5.0, 2.5.1 (if used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6) that may lead to shifts applied to segmentation objects in iPlan being not reflected when exported into DICOM under specific circumstances.
Source facts
- Action
- As of September 25, 2019 for new installations or updates/upgrades to the affected SW version, the Field Safety Notice / Product Notification letter is hand delivered to the customer during installation (before clinical use) by the according Brainlab Service representative All other consignees were notified by email on 10/04/2019. The letter details how the even occurs and provides instruction on how to avoid the issue until a software revision is disseminated.
- Initiated
- 2019-09-25
- Root cause
- Software design
- Terminated
- 2024-03-25
- Distribution
- worldwide
- Recall number
- Z-0176-2020
- Recall status
- Terminated
- CDRH record ID
- 176885
- FDA record created
- 2019-10-23
- Recall event number
- 83978