Z-0176-2022: ASC PneumoLiner device, Part No. WA90500US (US market only) The Pneumoliner device is intended for use a

Recall record Z-0176-2022 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0176-2022
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Oct 26, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing error, the Pneumoliner Bag Distal Tab that exits the Introducer shaft is in the wrong orientation. This results in the user deploying the bag upside down, which will make tissue encapsulation and bag closure more difficult, introducing the risk of trapping the small bowel/viscera in the bag at closure resulting in patient injury.

Source facts

Action
On September 21, 2021, the firm, Olympus, sent an "URGENT: MEDICAL DEVICE FIELD REMOVAL ACTION" letters to affected customers. Customers were instructed to do the following: 1. Immediately assess any affected product you have in stock and quarantine any affected product. 2. Call your Olympus customer service representative at 1-888-524-7266 option 1. Olympus will issue a Return Material Authorization in order to return any affected product at no charge to you. Olympus will issue a credit or replacement to your facility for your affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0401 and provide your contact information as indicated in the portal. In addition, if you may have further distributed this product, please identify your customers, notify them at once of this field removal action and appropriately document your notification process. If you require additional information, please contact Global V.P.at (484)896-5688 from Monday till Friday 8AM ET to 5PM ET or by e-mail at laura.storms@olympus.com.
Initiated
2021-09-21
Root cause
Nonconforming Material/Component
Terminated
2023-09-08
Distribution
US Nationwide distribution to states of: AL, CA, CT, FL, IL, MD, MI, MN, MO, NE, NY, OH, OK, OR, PA, SC, UT, VA, WA, WI, and WV ; and International (foreign) distribution to: Europe.
Recall number
Z-0176-2022
Recall status
Terminated
CDRH record ID
189439
FDA record created
2021-10-26
Recall event number
88750

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗