Open original FDA record ↗Recall record
- FDA identifier
- Z-0177-2019
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA record creation date
- Oct 18, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The terminals on the plug may become loose and overheat.
Source facts
- Action
- The firm, Siemens Healthineers, disseminated the " Customer Safety Advisory Notice" on 09/28/2018 to its customers. The notice described the product, problem and actions to be taken. The notice stated the potential problem and how to identify it. The customers were instructed to do the following: in the event overheating occurs or notice a distinct burning odor; stop use of your system and contact Siemens Healthineers service representatives.The issue will be addressed through a service visit to check if the terminal screws are torqued to the proper specification. If you have any questions, please contact your local Siemens Healthineers service representatives at the following numbers: America: 1-800-888-7436 Europe, Middle East, and Africa: +49 9131 940 4000 Asia and Australia: +86(21)3811 2121
- Initiated
- 2018-09-28
- Root cause
- Process control
- Terminated
- 2021-02-11
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Argentina, Austria, Australia, Azerbaijan, Belgium, Bulgaria, Brunnel, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Dominican Rep, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Republic Korea, Lebanon, Sri Lanka, Macedonia, Mexico, Malaysia, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Paraguay, Qatar, Romania, Russian Fed, Saudi Arabia, Singapore, Slovenia, Slovakia, Spain, South Africa, Switzerland, Thailand, Turkey, Taiwan, UAE, United Kingdom and Vietnam.
- Recall number
- Z-0177-2019
- Recall status
- Terminated
- CDRH record ID
- 167509
- FDA record created
- 2018-10-18
- Recall event number
- 81144