Skip to content

Public FDA record

Z-0177-2020: Minimed Model 500 Remote Control (MMT-500, MMT-500RU, MMT-500RUS, MMT-500RSE, MMT-500RCH, MMT-500RDE, MMT

Recall record Z-0177-2020 for Medtronic Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0177-2020
Company in FDA record
Medtronic Inc.
FDA classification date
Nov 4, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is a potential security vulnerability related to the use of the remote controller accessories with the insulin pumps.

Source facts

Action
Medtronic sent an Urgent Field Safety Notification letter dated August 7, 2018 to the patient, distributor, and healthcare provider via email, U.S. mail, or phone. The letter identified the affected product, problem and actions to be taken. Starting on 10/05/2021, recall notices were sent to patients, health care professionals, distributors, and personnel at clinical sites. Distributors were asked to provide their customer lists to the recalling firm so the recalling firm could send customers recall notices. The scope of the recall is being expanded to all users regardless of warranty status. Users are asked to stop using and disconnect the remote controller, disable the remote feature, and return the remote controller to the recalling firm. Customers were also asked to provide the information requested in the confirmation form to the recalling firm via phone, mail, email, or online on the firm's website. Additional questions can be directed to Technical Support at 1-800-378-2292. Disconnecting/Disabling instructions were provided in the recall notice, which provided the following link: https://info.medtronicdiabetes.com/remote-controller-return
Initiated
2018-08-07
Root cause
Software design
Terminated
2024-03-29
Distribution
Distribution was nationwide to consumers and distributors. There was no government/military distribution. Foreign distribution was made to Algeria, Australia, Austria, Belgium, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Estonia, Faroe Islands, Finland, France, Germany, Greece, Hungary, Iceland, India, Iran, Iraq, Ireland, Israel, Italy, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Martinique, Namibia, Netherlands, New Caledonia, Norway, Oman, Pakistan, Poland, Reunion, Romania, Russian Federation, Saudi Arabia, Senegal, Slovakia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, and United Kingdom.
Recall number
Z-0177-2020
Recall status
Terminated
CDRH record ID
170857
FDA classified
2019-11-04
Recall event number
82078

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗