Z-0178-2019: Biograph mCT: a) S(64)-3R, Material Number 10248669 b) S (20) -3R, Material Number 10507786 c) Flow 20-4

Recall record Z-0178-2019 for Siemens Medical Solutions USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0178-2019
Company in FDA record
Siemens Medical Solutions USA, Inc.
FDA record creation date
Oct 18, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The terminals on the plug may become loose and overheat.

Source facts

Action
The firm, Siemens Healthineers, disseminated the " Customer Safety Advisory Notice" on 09/28/2018 to its customers. The notice described the product, problem and actions to be taken. The notice stated the potential problem and how to identify it. The customers were instructed to do the following: in the event overheating occurs or notice a distinct burning odor; stop use of your system and contact Siemens Healthineers service representatives.The issue will be addressed through a service visit to check if the terminal screws are torqued to the proper specification. If you have any questions, please contact your local Siemens Healthineers service representatives at the following numbers: America: 1-800-888-7436 Europe, Middle East, and Africa: +49 9131 940 4000 Asia and Australia: +86(21)3811 2121
Initiated
2018-09-28
Root cause
Process control
Terminated
2021-02-11
Distribution
Worldwide Distribution: US (nationwide) and countries of: Argentina, Austria, Australia, Azerbaijan, Belgium, Bulgaria, Brunnel, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Dominican Rep, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Republic Korea, Lebanon, Sri Lanka, Macedonia, Mexico, Malaysia, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Paraguay, Qatar, Romania, Russian Fed, Saudi Arabia, Singapore, Slovenia, Slovakia, Spain, South Africa, Switzerland, Thailand, Turkey, Taiwan, UAE, United Kingdom and Vietnam.
Recall number
Z-0178-2019
Recall status
Terminated
CDRH record ID
167510
FDA record created
2018-10-18
Recall event number
81144

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗