Recall record Z-0178-2023 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0178-2023
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Nov 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.

Source facts

Action
The firm issued an Urgent Medical Device Correction notice to their consignee on 09/20/2022 by email. The notice explained the problem, risk, and requested the following: "Actions to be taken by Customer: 1. Identify the impacted lot number listed on page 1 and stop any use of the product immediately. 2. Hillrom will contact you to perform a correction of your impacted device as soon as possible." The firm issued an Urgent Medical Device Recall notice to their consignees on 09/27/2022 by FedEx 2nd day delivery. The notice explained the problem, risk, and requested the following: "Actions to be taken by Customer: 1. Identify the impacted lot numbers listed in the above table and replace with the enclosed device that accompanies this notification. 2. Please return the impacted unit per RMA instructions attached within 10 business days. Actions to be taken by Distributor: 1. Please forward this notification to your end users along with replacement parts accompanying this notification. 2. Please request return of the impacted devices from your end users and return the devices to Hillrom, per RMA instructions attached." Both notices provided the following contact information: "Contact reference person If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support at 800-445-3720 or technical.support@hillrom.com"
Initiated
2022-09-20
Root cause
Other
Terminated
2026-08-04
Distribution
MI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South Africa
Recall number
Z-0178-2023
Recall status
Terminated
CDRH record ID
196045
FDA classified
2022-11-03
Recall event number
90848

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗