Open original FDA record ↗Recall record
- FDA identifier
- Z-0179-06
- Company in FDA record
- FHC, Inc.
- FDA classification date
- Nov 23, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Power Assist microTargeting System may advance beyond intended target
Source facts
- Action
- FHC Inc. and Medtronic notified customers by letter and instructs the customer to return their Motor Option for upgrade to include the following components: 1) Handheld Remote Control, 2) Motor Controller Unit with Display, 3) Motor, and 4) Medical Power Adapter. (See pictures at the end of this communication). Until the Motor Option has been updated, instructions to customers are: Prior to each clinical use: Ensure the microTargeting Drive system is calibrated before each clinical use per the User Guide. Check the microTargeting Drive system functionality over a range of at least 10mm movement. Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. During clinical use: Should any performance indicating either a stall or unintended movement be observed, immediately unplug the 12V Medical Power Adapter, remove the MicroTargeting Drive Motor Option from the system, and continue using the microTargeting Drive in the manual mode. Every 100 uses or annually: Have your microTargeting Drive system checked by the manufacturer per the User Guide. Have your equipment support department check or tighten the MicroTargeting Drive Motor Option Remote Control Knob set screws, tighten as needed, and then re-calibrate.
- Initiated
- 2005-10-31
- Root cause
- Other
- Terminated
- 2006-06-07
- Distribution
- Nationwide Foreign: South Africa, Taiwan, Hong Kong, China, France
- Recall number
- Z-0179-06
- Recall status
- Terminated
- CDRH record ID
- 42773
- FDA classified
- 2005-11-23
- Recall event number
- 33946