Z-0179-2007: ConMed Linvatec Sternum Saw Collet Nut, Catalog Number 5059-09: Used on ConMed Linvatec Sternum Saw Handp

Recall record Z-0179-2007 for Linvatec Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0179-2007
Company in FDA record
Linvatec Corp.
FDA classification date
Nov 16, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The Sternum Saw Collet Nut that is used to insert and hold the saw blade in place may not always allow for the easy insertion of a saw blade into the handpiece for use. The nut needs to be in a certain position in order for the blade to be inserted and this may delay the operation of the device especially in emergency situations.

Source facts

Action
Linvatec will send each consignee a recall package that will include: a 12/20/2005 recall letter and reply form, A product correction instruction sheet , a new Sternum Saw Collet Nut and an assembly manual. Linvatec will also contact each consignee by phone or email. All consignees that have not responded by 30 days after the first mailing will be re-contacted by the firm by mail, phone or email.
Initiated
2005-12-20
Root cause
Other
Terminated
2012-07-11
Distribution
Worldwide, including USA, Mexico, and Canada.
Recall number
Z-0179-2007
Recall status
Terminated
CDRH record ID
43846
FDA classified
2006-11-16
Recall event number
34438

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗