Open original FDA record ↗Recall record
- FDA identifier
- Z-0179-2007
- Company in FDA record
- Linvatec Corp.
- FDA classification date
- Nov 16, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Sternum Saw Collet Nut that is used to insert and hold the saw blade in place may not always allow for the easy insertion of a saw blade into the handpiece for use. The nut needs to be in a certain position in order for the blade to be inserted and this may delay the operation of the device especially in emergency situations.
Source facts
- Action
- Linvatec will send each consignee a recall package that will include: a 12/20/2005 recall letter and reply form, A product correction instruction sheet , a new Sternum Saw Collet Nut and an assembly manual. Linvatec will also contact each consignee by phone or email. All consignees that have not responded by 30 days after the first mailing will be re-contacted by the firm by mail, phone or email.
- Initiated
- 2005-12-20
- Root cause
- Other
- Terminated
- 2012-07-11
- Distribution
- Worldwide, including USA, Mexico, and Canada.
- Recall number
- Z-0179-2007
- Recall status
- Terminated
- CDRH record ID
- 43846
- FDA classified
- 2006-11-16
- Recall event number
- 34438