Reports of the Vacuette QuickShield Safety Holders breaking during use due to a crack at the holder hub.
Source facts
Action
Greiner Bio-One sent a "QUICKSHIELD WITH SAFETY HOLDER" notification letter dated August 11, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to stop distributing and using the affected products. A Product Disposition Form was attached for customers to complete and return for replacements. Contact the firm at 888-266-3883 for questions concerning this recall.
Initiated
2011-08-11
Root cause
Nonconforming Material/Component
Terminated
2012-03-20
Distribution
Nationwide Distribution-USA (nationwide) including the states of CO, IL, KY, NY, PA and VA.
Recall number
Z-0179-2012
Recall status
Terminated
CDRH record ID
103997
FDA classified
2011-11-09
Recall event number
59918