Open original FDA record ↗Recall record
- FDA identifier
- Z-0180-2010
- Company in FDA record
- Boston Scientific CRM Corp
- FDA classification date
- Nov 12, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The L ATITUDE communicator (Model 6481) for a patient may not have been able to receive measurements from their weight scale or blood pressure monitor afer initial setup. The issue has been resolved. If any patient weight or blood pressure readings are stored in the patient's sensor, they will be uploaded to LATITUDE after the patient's communicator calls in to Boston Scientific's servers. It
Source facts
- Action
- Boston Scientific used phone calls as the delivery method for these communications. The communication began August 6, 2009 for the original 6 patients affected. Notifications via phone calls to additional patients affected began on August 17, 2009. Direct questions about the recall to the LATITUDE Customer Service by calling 1-651-582-4000.
- Initiated
- 2009-08-07
- Root cause
- Software design
- Terminated
- 2012-12-04
- Distribution
- Nationwide Distribution -- AZ, AR, CA, CT, FL, GA, IL, IN, MA, MO, NV, NJ, NY, OH, OR, PA, SC, TX, UT, VA and WA.
- Recall number
- Z-0180-2010
- Recall status
- Terminated
- CDRH record ID
- 84609
- FDA classified
- 2009-11-12
- Recall event number
- 53083