Covidien is recalling certain Puritan Bennett 980 Ventilator Systems due to a blacklight inverter PCBA where a capacitor component can fail, causing a dim but functional screen and thermal odor.
Source facts
Action
Update: Consignees were notified on 12/11/14 of strategy change after Covidien sent an Urgent Field Safety Corrective Action notification with instructions dated September 2014 via traceable courier to affected consignees. The letter identifed the affected product, problem and actions to be taken. Consignees were instructed to complete the attached verification form and fax it to the Covidien RMS contact indicated on the form. If you need further assistance, contact the Technical Support Department @ 1-800-255-6774, option 4, then option 1.
Initiated
2014-10-01
Root cause
Device Design
Terminated
2016-07-01
Distribution
Worldwide Distribution-US (nationwide) including the states of CA, CO, FL, GA, KY, MA, MN, NC, NY, OH, OK, PA, SC, TN, TX, UT, and WI, and the countries of Canada, Mexico, Saudi Arabia, South Africa and United Arab Emirates.
Recall number
Z-0180-2015
Recall status
Terminated
CDRH record ID
130229
FDA classified
2014-12-09
Recall event number
69324