Open original FDA record ↗Recall record
- FDA identifier
- Z-0180-2016
- Company in FDA record
- Encore Medical, Lp
- FDA record creation date
- Oct 30, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
Source facts
- Action
- DJO Global sent an Urgent Field Safety Notice dated September 15, 2015, to all affected consignees. The notice identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed become familiar with the new techniques and notify all of their applicable users of the RSP Monoblock and Turon stems of these revised methods. When each user has been notified of these new techniques, they should complete the form that was provided with the notification indicating that they have been informed of these changes and understand the new methods. Consignees were asked to complete and return forms. Consignees with questions should call 512-834-6255.
- Initiated
- 2015-09-14
- Root cause
- Device Design
- Terminated
- 2016-11-04
- Distribution
- Nationwide Distribution
- Recall number
- Z-0180-2016
- Recall status
- Terminated
- CDRH record ID
- 140477
- FDA record created
- 2015-10-30
- Recall event number
- 72229