- Recall classification
- Class II
Reason stated in the source
Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.
Source facts
- Action
- Philips Medical Systems issued certified letters on 7/8/14 and receipt returned for tracking purposes. A mandatory Field Change Order with reference FC072200253 is being released that requires Philips field service engineers to install Software release R8.2.0.3. If you need any further information or support concerning this issue, contact your local Philips representative: SUPPORT 1-800-722-9377.
- Initiated
- 2014-07-08
- Root cause
- Software design
- Terminated
- 2017-01-18
- Distribution
- Distributed in the states of IL, and TN, and the countries of Argentina , Australia, Belgium, Denmark, Egypt, France, Germany, Hong Kong, Italy, Japan, Korea, Republic of Netherlands, Norway, Saudi Arabia, Switzerland, United Kingdom, and Viet Nam.
- Recall number
- Z-0181-2015
- Recall status
- Terminated
- CDRH record ID
- 128567
- FDA classified
- 2014-11-07
- Recall event number
- 68723